Transorbital B-mode ultrasound for the assessment of posterior globe flattening in idiopathic intracranial hypertension: a pilot study

Abstract

Background

Posterior globe flattening (PGF) is a specific neuroimaging sign in patients with idiopathic intracranial hypertension (IIH), but its detection is based on subjective qualitative neuroradiological assessment. This study sought to evaluate the utility of transorbital ultrasound to detect and quantify PGF in IIH patients using the Posterior Globe Angle (PGA).

Methods

Consecutive IIH patients and healthy controls were enrolled in a prospective case-control study. Transorbital ultrasound was performed to assess the presence of PGF. For quantification of PGF, an angular measurement (PGA) was performed with the vertex centering the optic nerve at a predefined distance from the lamina cribrosa and angle legs tangentially aligned to the borders of the vitreous body. PGA measurements were compared between IIH patients and healthy controls. Additionally, the diagnostic accuracy of PGA measurements in detecting PGF was evaluated using ROC analysis.

Results

Thirty-one IIH patients (37.3 ± 12.3 years, 29 female) and 28 controls (33.3 ± 11.8 years, 21 female) were compared. PGF was present in 39% of IIH patients and absent in the control group. PGA3mm measurements significantly differed between IIH and controls (116.5° ± 5.5 vs. 111.7° ± 2.9; p < 0.001). A PGA3mm cutoff of ≥ 118.5° distinguished IIH patients from controls with 100% specificity, while retaining a sensitivity of 37.5%.

Conclusions

Transorbital ultrasound may be applied to detect and quantify PGF in IIH patients. Prospective, multicenter studies with extended cohorts and blinded design are needed to validate these preliminary findings and confirm the diagnostic utility of transorbital ultrasound for the assessment of PGF in IIH.

Introduction

Idiopathic intracranial hypertension (IIH) is a rare headache syndrome mainly observed in obese women of reproductive age [1,2,3]. The condition is defined by the presence of an elevated intracranial pressure without identifiable source of hydrocephalus or cerebral mass lesions. Its underlying pathophysiology remains vague [4]. Principal symptoms are headaches, visual disturbance and diplopia [5]. In a retrospective review of 165 patients referred to the neuro-ophthalmology division of a tertiary care center due to suspected IIH, Fisayo et al. found a misdiagnosis rate of 39.5%, emphasizing the importance of a thorough diagnostic work-up and the application of current IIH diagnostic criteria [6]. According to the revised Friedman criteria a diagnosis of IIH can be established in patients with papilledema and elevated lumbar puncture opening pressure (≥ 25cmCSF) if alternate diagnoses have been excluded through adequate neurological examination, neuroimaging, and cerebrospinal fluid (CSF) analysis (Table 1) [7]. However, the Friedman criteria also allow for a diagnosis of “possible IIH” in patients without papilledema, if 3 of 4 characteristic neuroimaging criteria are present. Magnetic resonance imaging (MRI) features that are suggestive of IIH include the presence of (1) an “empty sella”, (2) distension of the optic nerve sheath, (3) transverse sinus stenosis and (4) posterior globe flattening (PGF).

Table 1 IIH diagnostic criteria. Table modified according to Friedman et al. [7]

PGF describes a flattening of the posterior part of the ocular globe, which has primarily been associated with IIH and with secondary intracranial hypertension. It is considered to result from elevated CSF pressure being transmitted from the subarachnoid space through the optic nerve sheath to the posterior globe [8]. Interestingly, PGF has also been reported in astronauts as a part of spaceflight associated neuro-ocular syndrome [9]. Various studies have found PGF to be among the most specific imaging signs of IIH (range 78–100%, [10,11,12]), whereas reports on the sensitivity of PGF in IIH vary considerably (44–80%, [10, 13]), possibly resulting from uncertainties in determining the presence of PGF. Currently, the assessment for PGF is based on a subjective radiological review of the available brain MRI. As of today, there are no standardized imaging protocols for the assessment of PGF. Subsequently, PGF is either documented as present or not. Quantitative methods are not routinely utilized. However, the quality of orbital MRI may vary depending on the burden of technical artifacts (e.g. motion artifacts, partial volume-, susceptibility-, or chemical shift artifacts) and on the scanning protocol used for image acquisition (e.g. slice thickness and field strength) [14].

Transorbital ultrasound allows for a non-invasive, dynamic, and high-resolution examination of the posterior aspect of the globe. In IIH, ultrasound is routinely applied for the measurement of the optic nerve sheath diameter (ONSD) as an indicator of elevated intracranial pressure [15]. Additionally, ultrasound has shown utility in the quantitative assessment of papilledema (by measuring the protrusion of the optic nerve head into the globe) and the differentiation of pseudo-papilledema due to the presence of optic disc drusen [16,17,18]. To the best of our knowledge, ultrasound has not yet been applied to assess PGF in IIH.

This study aimed (1) to evaluate the feasibility of ultrasound imaging of PGF and (2) to develop a quantitative assessment protocol for PGF in IIH and healthy controls using B-mode transorbital ultrasound.

Methods

This prospective case-control study evaluated PGF on transorbital ultrasound of the globes of IIH patients and matched healthy controls. All subjects gave informed, written consent to participate in this study. Ethical approval was obtained and the study was registered with the German registry for clinical trials (DRKS) (ID-No. DRKS00017815).

Study population

This study included consecutive adult patients who underwent diagnostic evaluation or inpatient treatment for IIH between January 2021 and November 2022 at our tertiary care neurological department. Patients were considered for study inclusion if (1) a diagnosis of IIH had been established prior to current admission or (2) a diagnosis of IIH was suspected. Consecutively, diagnosis of IIH was reviewed by the study team according to the revised Friedman criteria [7], based on ophthalmological and neurological findings, cerebral magnetic resonance imaging and measurements of cerebrospinal fluid opening pressure. Patients were excluded from the study if IIH diagnostic criteria were not met and/or a secondary cause of intracranial hypertension was identified. Patients without fundoscopic evidence of papilledema (current or in medical history) were not included in this study. Figure 1 depicts the patient selection process.

Fig. 1
figure 1

Flow-chart depicting patient selection. Forty-three patients were considered for study inclusion. The IIH diagnosis was reviewed according to the revised Friedman criteria [7]. Patients were excluded from the study if IIH diagnostic criteria were not met and/or a secondary cause of intracranial hypertension was identified

Healthy subjects, free of current or past neurological or ophthalmological disorders, matched for age and sex were recruited as controls between January 2021 and February 2023.

Transorbital ultrasound

Ultrasound examinations were performed by two experienced investigators (LD and TK) using a 7.5-MHz linear-array transducer of the Aplio 300 ultrasound system (Canon Medical Systems Corp., Ōtawara, Tochigi, Japan) and an orbital imaging preset, limiting the acoustic output exposure levels according to the recommendations of the U.S. Food and Drug Administration [19]. Both investigators were aware of the participants medical records and IIH diagnosis at the time of the examination. Care was taken to limit the ultrasound exposure (dwell time) to less than a minute for each eye. Standard B-mode transorbital ultrasound in the axial plane included measurements of optic nerve diameter (OND), optic nerve sheath diameter (ONSD) and optic disc size. If ONSD and optic disc size differed between a patients’ eyes, the most severely affected eye (i.e. the higher value) was used for subsequent group comparison. OND values are reported as the mean value of both eyes.

For the qualitative assessment of the presence of PGF, the acquired B-mode images of each orbit were reviewed by TK and LD. Posterior globe flattening was considered present if there was a straightening of the normal outward convexity of the sclera at the area of attachment to the optic nerve in at least one eye (Fig. 2).

Fig. 2
figure 2

Two B-mode transorbital sonographic images of two individual globes in the axial plane. Image A shows a normal globe. Image B depicts posterior globe flattening in a globe with straightening of the outward convexity of the sclera at the area of attachment to the optic nerve (yellow arrowheads)

For the quantitative assessment of PGF, a transverse imaging plane showing the insertion of the optic nerve into the eyeball was set first. Next, the vertex for angular measurement was marked at the center of the optic nerve at a predefined distance from the lamina cribrosa, at 2.0 mm, 2.5 mm and 3.0 mm, respectively. Starting from the vertex, the two angle legs were tangentially aligned with the medial and lateral border of the anechoic vitreous body. Subsequently, the Posterior globe angle (PGA = α) was measured between the angle legs (Figs. 3, 4 and 5). If the PGA differed between a patients’ eyes, the most severely affected eye (i.e. the higher value) was used for statistical analysis.

Fig. 3
figure 3

Graphic illustration of the PGA (α) measured at 3.0 mm distance from the ophthalmic lamina cribrosa

Fig. 4
figure 4

Transorbital ultrasound of one globe in the axial plane. Picture A shows the original image, the yellow arrowhead marks the ophthalmic lamina cribrosa. Picture B shows exemplary measurement of the PGA at 3.0 mm distance from the ophthalmic lamina cribrosa

Fig. 5
figure 5

Transorbital ultrasound images of one globe in the axial plane depicting measurements of the PGA at 3.0 mm distance from the ophthalmic lamina cribrosa in a globe with PGF (A) and a normal globe (B)

Fig. 6
figure 6

ROC curves of the PGA values obtained by transorbital ultrasound to differentiate the IIH patients from healthy controls. PGA was measured at 3.0 mm, 2.5 mm and 2.0 mm distance from the ophthalmic lamina cribrosa. Corresponding data is summarized in Table 4

Statistical analysis

Statistical analyses were performed using IBM SPSS Statistics software (IBM SPSS Statistics, Version 29.0. Armonk, NY: IBM Corp.). The continuous variables are presented as mean ± standard deviation.

Given that the Q-Q plots and histograms indicated normal distribution for all continuous variables, two-tailed t-tests were employed for comparing groups across these variables. Two-sided 95% confidence intervals are provided for the differences in means. p-values were adjusted for multiple comparisons using the Bonferroni-Holm procedure, where applicable. Group comparisons concerning sex were computed using Fisher’s exact test. Sensitivity and specificity were calculated by utilizing receiver operating characteristic (ROC) curves and the Youden index for cut-point estimation.

Results

Study population

During the study period 43 patients were considered for study inclusion. Among them, 25 patients were referred with suspected IIH, while 18 patients had been diagnosed with IIH prior to current admission. Twelve patients did not meet the diagnostic criteria for IIH and were excluded for either (1) lack of papilledema (n = 8), and/or (2) lack of elevated CSF opening pressure (n = 3) or (3) due to the identification of sinus venous thrombosis (n = 1).

The final study cohort consisted of 31 consecutive IIH patients (29 female; 37.3 ± 12.3 years) and 28 healthy controls (21 female; 33.3 ± 11.8 years). Patients and controls did not significantly differ regarding age (p = 0.200) and sex (p = 0.071). The IIH cohort included patients with active IIH as determined by presence of papilledema, and patients in ocular remission (i.e. no current evidence of papilledema). When they participated in study, 65% (20/31) of the IIH cohort had papilledema, 13% (4/31) had optic nerve atrophy and 23% (7/31) had no evidence of papilledema anymore. The mean CSF opening pressure before measurement of PGF was 36 ± 10.5 cmCSF (range: 17–55 cmCSF). Table 2 shows the characteristics of the study population.

Table 2 Characteristics of the study population

Transorbital ultrasound

Table 3 summarizes transorbital ultrasound measurements in IIH patients and controls. OND (2.6 ± 0.5 mm vs. 2.9 ± 0.5 mm, p = 0.031) and ONSD (6.0 ± 0.7 mm vs. 5.1 ± 0.5 mm, p < 0.001) significantly differed between IIH patients and controls. No significant differences were observed for optic disc measurements (patients: 0.7 ± 0.5 mm, controls: 0.5 ± 0.9 mm, p = 0.419, 95% CI: -0.6–0.3 mm).

Table 3 Results of the transorbital ultrasound. Values are presented as mean ± standard deviation. *Bonferroni-Holm corrected values

Qualitative assessment of the globes found PGF to be present on transorbital ultrasound in 39% of IIH patients. None of the control subjects exhibited PGF. The PGA measured at 3.0 mm, 2.5 mm and 2.0 mm distance from the lamina cribrosa all significantly differed between IIH patients and the control group (Table 3). The PGA at 3.0 mm showed the highest difference of the means between IIH patients and controls (95% CI: 2.5–7.2°, p < 0.001). Similarly, in ROC analysis the PGA at 3.0 mm yielded the highest area under the curve (AUC = 0.74) when compared with PGA measured at 2.5 mm and 2.0 mm distance from the lamina cribrosa (Table 4, Fig. 6). A predefined cutoff ≥ 118.5° PGA3mm identified IIH patients with a specificity of 100%, retaining a sensitivity of 37,5% (Youden index: 0.375). Table 4 summarizes sensitivity and specificity obtained by ROC analysis and optimal cutoff points estimated by Youden index for PGA measurements at 3.0 mm, 2.5 mm and 2.0 mm distance from the lamina cribrosa.

Table 4 Sensitivity and specificity analysis for the PGA measured at 3.0 mm, 2.5 mm and 2.0 mm distance from the lamina cribrosa. Area under the curve (AUC), sensitivity, and specificity were obtained via ROC analysis. ROC curves are depicted in Fig. 6. Optimal cutoff points were estimated using the Youden index. The maximum value of the Youden index is 1 (perfect test) and the minimum is 0

Discussion

This study applied transorbital ultrasound to quantify PGF in patients with IIH using the posterior globe angle (PGA). Here, we were able to show for the first time, that PGA measurements differ significantly between IIH patients and healthy controls. We compared PGA measurements at different positions of the angle vertex using ROC analysis, which found the highest AUC for measurements at 3.0 mm distance from the lamina cribrosa (AUC = 0.74). A cutoff of ≥ 118.5° PGA would identify IIH patients with 100% specificity while retaining a sensitivity of 37.5%. The results of this preliminary case-control study indicate that transorbital ultrasound may be applied to identify and quantify PGF in IIH patients.

In 1993, Gibby et al. described a “flattening of [the] normal outward convexity” of the posterior sclera on computed tomography images of IIH patients [20]. Since then, multiple studies have investigated PGF in IIH using magnetic resonance imaging [10, 11, 13, 21,22,23,24,25,26]. The overwhelming majority of these studies based the radiological assessment on a qualitative, subjective evaluation of axial T2-weighted images. Consequently, reports on diagnostic accuracy have produced inconsistent results, with sensitivity and specificity of PGF in IIH varying between 28.8 and 78.6% and 59.5–99.1%, respectively [27]. Similarly, reports by Delen et al. and Alperin et al. only show moderate interrater-reliability of quantitative PGF evaluation (k = 0.46–0.64) [22, 28]. Still, the available studies suggest that PGF may show diagnostic utility as a rather specific imaging finding.

In 2013, Alperin et al. reported the first quantitative assessment of PGF and optic nerve protrusion in IIH using MRI in a cohort of 7 IIH patients and 6 healthy controls [28]. The authors developed software to convert the three-dimensional structure of the globe into a color-coded two-dimensional distance map, visualizing the distances between the center of the globe to different points on the posterior sclera. While Alperin et al. argued automated, quantitative measurements would likely improve the detection of globe distortions in IIH, their findings have yet to be confirmed in subsequent studies using larger cohorts.

Compared to MRI, transorbital ultrasound is a cost-effective, widely available, and easy-to-use examination method that is particularly suitable for follow-up examinations. Since it is already established to measure OND and papilledema in routine clinical practice, it seems reasonable to extend its scope of application to assess deformations of the posterior ocular globe in IIH. The diagnosis of IIH remains a challenge, as reports on diagnostic error are prominent in the medical literature [6, 29]. The improvement of available diagnostic modalities, as well as the establishment of new diagnostic approaches could improve detection and disease monitoring of IIH.

This preliminary study is limited by its small case size due to the rarity of the disease. Although the current diagnostic criteria were applied, the IIH cohort included patients with active IIH and patients in the stage of ocular remission, which may have affected our results. In addition, the examiners who performed the ultrasound were not blinded to the patients’ medical history, including IIH diagnosis, which introduces a risk of bias. Consequently, further studies are necessary to confirm our findings using extended cohorts, a blinded study design and concomitant MR-imaging of the ocular globes as reference.

It is currently unclear to what extent PGA measurements are influenced by morphological variations of the ocular globe, e.g. in myopia or hyperopia. Theoretically, transorbital ultrasound examination could be extended to include further measurements to adjust for differences in ocular globe morphology. For PGA measurements, a transverse image plane at the insertion of the optic nerve into the eyeball is set. However, bulb movements may impede the assessment and lead to oblique imaging planes, possibly distorting measurements. We recommend placing the examiner’s proximal wrist on the patient’s forehead to avoid unnecessary pressure on the eye during the assessment and to ensure sufficient application of ultrasound gel. These precautions, in addition to direct instruction of the patient, help to reduce the occurrence of bulbar movements.

Conclusion

To conclude, this prospective case-control study is the first to apply transorbital ultrasound to depict and to quantify PGF in IIH. PGA measurements differed significantly between IIH patients and controls. Measured at 3.0 mm distance from the lamina cribrosa, a predefined PGA cut-off of ≥ 118.5° yielded a sensitivity of 37.5% and specificity of 100% to distinguish IIH patients from controls. However, given the limitations of this study, its results should be considered preliminary. Prospective, multicenter studies with extended cohorts and blinded design are needed to confirm the diagnostic utility and reproducibility of transorbital ultrasound for the assessment of PGF in IIH.

Data availability

The datasets generated for the current study are not publicly available but are available from the corresponding author on reasonable request.

Abbreviations

IIH:

Idiopathic intracranial hypertension

PGF:

Posterior globe flattening

PGA:

Posterior globe angle

MRI:

Magnetic resonance imaging

OND:

Optic nerve diameter

ONSD:

Optic nerve sheath diameter

CSF:

Cerebrospinal fluid

ROC:

Receiver operating characteristic

AUC:

Area under the curve

References

  1. Balcer LJ, Liu GT, Forman S et al (1999) Idiopathic intracranial hypertension: relation of age and obesity in children. Neurology 52. https://doi.org/10.1212/WNL.52.4.870. 870 LP – 870

  2. Durcan PJ, Corbett JJ, Wall M (1988) The incidence of Pseudotumor Cerebri: Population studies in Iowa and Louisiana. Arch Neurol 45:875–877. https://doi.org/10.1001/archneur.1988.00520320065016

    Article  CAS  PubMed  Google Scholar 

  3. Galvin JA, Van Stavern GP (2004) Clinical characterization of idiopathic intracranial hypertension at the Detroit Medical Center. J Neurol Sci 223:157–160. https://doi.org/10.1016/j.jns.2004.05.009

    Article  PubMed  Google Scholar 

  4. Markey KA, Mollan SP, Jensen RH, Sinclair AJ (2016) Understanding idiopathic intracranial hypertension: mechanisms, management, and future directions. Lancet Neurol 15:78–91. https://doi.org/10.1016/S1474-4422(15)00298-7

    Article  PubMed  Google Scholar 

  5. Ball AK, Clarke CE (2006) Idiopathic intracranial hypertension. Lancet Neurol 5:433–442. https://doi.org/10.1016/S1474-4422(06)70442-2

    Article  PubMed  Google Scholar 

  6. Fisayo A, Bruce BB, Newman NJ, Biousse V (2016) Overdiagnosis of idiopathic intracranial hypertension. Neurology 86:341. https://doi.org/10.1212/WNL.0000000000002318

    Article  PubMed  PubMed Central  Google Scholar 

  7. Friedman DI, Liu GT, Digre KB (2013) Revised diagnostic criteria for the pseudotumor cerebri syndrome in adults and children. Neurology 81:1159–1165. https://doi.org/10.1212/WNL.0b013e3182a55f17

    Article  PubMed  Google Scholar 

  8. Degnan AJ, Levy LM (2011) Pseudotumor Cerebri: brief review of clinical syndrome and imaging findings. AJNR Am J Neuroradiol 32:1986–1993. https://doi.org/10.3174/AJNR.A2404

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Nelson ES, Mulugeta L, Myers JG (2014) Microgravity-induced fluid shift and ophthalmic changes. Life 4:621–665. https://doi.org/10.3390/life4040621

    Article  PubMed  PubMed Central  Google Scholar 

  10. Maralani PJ, Hassanlou M, Torres C et al (2012) Accuracy of brain imaging in the diagnosis of idiopathic intracranial hypertension. Clin Radiol 67:656–663. https://doi.org/10.1016/j.crad.2011.12.002

    Article  CAS  PubMed  Google Scholar 

  11. Agid R, Farb RI, Willinsky RA et al (2006) Idiopathic intracranial hypertension: the validity of cross-sectional neuroimaging signs. Neuroradiology 48:521–527. https://doi.org/10.1007/s00234-006-0095-y

    Article  CAS  PubMed  Google Scholar 

  12. Rohr A, Dörner L, Stingele R et al (2007) Reversibility of venous sinus obstruction in idiopathic intracranial hypertension. Am J Neuroradiol 28:656–659

    CAS  PubMed  PubMed Central  Google Scholar 

  13. Brodsky M (1998) Magnetic resonance imaging in pseudotumor cerebri. Ophthalmology 105:1686–1693. https://doi.org/10.1016/S0161-6420(98)99039-X

    Article  CAS  PubMed  Google Scholar 

  14. Ann JH, Kim EY (2015) Assessment of posterior Globe Flattening: two-dimensional versus three-dimensional T2-Weighted imaging. Invest Magn Reson Imaging 19:178. https://doi.org/10.13104/imri.2015.19.3.178

    Article  Google Scholar 

  15. Bäuerle J, Nedelmann M (2011) Sonographic assessment of the optic nerve sheath in idiopathic intracranial hypertension. J Neurol 258:2014–2019. https://doi.org/10.1007/S00415-011-6059-0/FIGURES/4

    Article  PubMed  Google Scholar 

  16. Tamburrelli C, Salgarello T, Grazia Caputo C et al (2000) Ultrasonographic evaluation of optic disc swelling: comparison with CSLO in idiopathic intracranial hypertension. Invest Ophthalmol Visual Sci 41:2960–2966

    CAS  Google Scholar 

  17. Neudorfer M, Ben-Haim MS, Leibovitch I, Kesler A (2013) The efficacy of optic nerve ultrasonography for differentiating papilloedema from pseudopapilloedema in eyes with swollen optic discs. Acta Ophthalmol 91:376–380. https://doi.org/10.1111/J.1755-3768.2011.02253.X

    Article  PubMed  Google Scholar 

  18. Carter SB, Pistilli M, Livingston KG et al (2014) The role of orbital ultrasonography in distinguishing papilledema from pseudopapilledema. Eye 28:1425–1430. https://doi.org/10.1038/EYE.2014.210

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  19. FDA (2023) Marketing Clearance of Diagnostic Ultrasound Systems and Transducers

  20. Gibby WA, Cohen MS, Goldberg HI, Sergott RC (1993) Pseudotumor Cerebri: CT findings and correlation with vision loss. Am J Roentgenol 160:143–146. https://doi.org/10.2214/ajr.160.1.8416612

    Article  CAS  Google Scholar 

  21. Bsteh G, Marik W, Krajnc N et al (2023) MRI features of idiopathic intracranial hypertension are not prognostic of visual and headache outcome. J Headache Pain 24:1–8. https://doi.org/10.1186/s10194-023-01641-x

    Article  Google Scholar 

  22. Delen F, Peker E, Onay M et al (2019) The significance and reliability of imaging findings in Pseudotumor Cerebri. Neuro-Ophthalmology 43:81–90. https://doi.org/10.1080/01658107.2018.1493514

    Article  PubMed  Google Scholar 

  23. Görkem SB, Doğanay S, Canpolat M et al (2015) MR imaging findings in children with pseudotumor cerebri and comparison with healthy controls. Child’s Nerv Syst 31:373–380. https://doi.org/10.1007/s00381-014-2579-0

    Article  Google Scholar 

  24. Hoffmann J, Huppertz HJ, Schmidt C et al (2013) Morphometric and volumetric MRI changes in idiopathic intracranial hypertension. Cephalalgia 33:1075–1084. https://doi.org/10.1177/0333102413484095

    Article  PubMed  Google Scholar 

  25. Butros SR, Goncalves LF, Thompson D et al (2012) Imaging features of idiopathic intracranial hypertension, including a new finding: Widening of the foramen ovale. Acta Radiol 53:682–688. https://doi.org/10.1258/ar.2012.110705

    Article  PubMed  Google Scholar 

  26. Lim MJ, Pushparajah K, Jan W et al (2010) Magnetic resonance imaging changes in idiopathic intracranial hypertension in children. J Child Neurol 25:294–299. https://doi.org/10.1177/0883073809338874

    Article  PubMed  Google Scholar 

  27. Kwee RM, Kwee TC (2019) Systematic review and meta-analysis of MRI signs for diagnosis of idiopathic intracranial hypertension. Eur J Radiol 116:106–115. https://doi.org/10.1016/j.ejrad.2019.04.023

    Article  PubMed  Google Scholar 

  28. Alperin N, Bagci AM, Lam BL, Sklar E (2013) Automated quantitation of the posterior scleral flattening and optic nerve protrusion by MRI in idiopathic intracranial hypertension. Am J Neuroradiol 34:2354–2359. https://doi.org/10.3174/ajnr.A3600

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  29. Krishnakumar D, Pickard JD, Czosnyka Z et al (2014) Idiopathic intracranial hypertension in childhood: pitfalls in diagnosis. Dev Med Child Neurol 56:749–755. https://doi.org/10.1111/dmcn.12475

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

LD is participant in the BIH - Charité Clinician Scientist Program funded by the Charité - Universitätsmedizin Berlin and the Berlin Institute of Health.

Funding

Open Access funding enabled and organized by Projekt DEAL. We acknowledge financial support from the Open Access Publication Fund of Charité – Universitätsmedizin Berlin and the German Research Foundation (DFG).

Open Access funding enabled and organized by Projekt DEAL.

Author information

Authors and Affiliations

Authors

Contributions

TK oversaw study design and conception. Data collection was performed by TK and LD. Data analysis and visualization was performed by TK and LD with counselling for statistical analysis by KN. CP and MRZ gave critical input regarding study design and conception. The draft of the manuscript was written by TK and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript and gave consent for publication.

Corresponding author

Correspondence to Theresia Knoche.

Ethics declarations

Ethics approval and consent to participate

This research study was conducted in approval with the ethics committee of Charité – Universitätsmedizin Berlin (Application No. EA2/088/19) and the study was registered with the German Registry for clinical trials (DRKS) (ID-No. DRKS00017815). All participants gave written informed consent to participate in this study and for publication of their anonymized data.

Conflict of interest

The authors declare that they have no financial or non-financial conflicts of interest to disclose. The authors are indebted to the patients for their participation in this study.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions