Intra- and inter-rater reliability of rectus femoris muscle thickness measured using ultrasonography in healthy individuals

Abstract

Background

Ultrasonography (US) is a feasible and accessible method for the measurement of skeletal muscle mass. This technique presents acceptable intra-rater reliability; however, there are a few reports on its inter-rater reliability. Additionally, relative reliability should equally be inspected to determine the presence of systematic errors. Therefore, this study aimed to investigate the intra- and inter-rater reliabilities and absolute reliability of rectus femoris muscle thickness as measured using US.

Methods

The participants included in our study comprised 12 healthy young men (26.5 ± 3.9 years. Rectus femoris muscle thickness was measured from the right side of the thigh using US by two trained physical therapists. Inter- and intra-rater reliabilities were determined using the intraclass correlation coefficient (ICC) (1, 1) and ICC (2, 1) methods, respectively. Absolute reliability was evaluated using Bland − Altman analysis. Additionally, we calculated the minimal detectable change at the 95% level of confidence (MDC95).

Result

According to the results of the Bland − Altman analysis, no fixed or proportional errors were present. The ICC (1, 1) was 0.95, and the ICC (2, 1) was 0.70. The MDC95 values of rectus femoris thickness for the intra- and inter-rater reliabilities were 2.0 mm and 4.3 mm, respectively.

Conclusions

In our study, intra- and inter-rater reliabilities were measured at “excellent” and “moderate” levels in the healthy individuals based on a previously defined scale. Moreover, we determined the measurement error for quantifying rectus femoris thickness. Therefore, the measurement of rectus femoris thickness using US could be considered applicable in clinical research.

Background

Loss of skeletal muscle mass is a major and well-known quantitative change associated with aging. Similarly, qualitative changes in skeletal muscle have been reported to correlate with aging, such as the preferential atrophy of type II muscle fibers [1, 2], increased intramuscular fat [3,4,5], and increased extracellular water volume relative to muscle volume [6]. The gold standard methods for skeletal muscle mass measurement are computed tomography (CT), magnetic resonance imaging (MRI), and dual-energy X-ray absorptiometry (DEXA) [7]. However, these techniques are not optimal for measuring qualitative changes in the skeletal muscle. Furthermore, they are not frequently used as they are costly and infer a risk of radiation exposure in a clinical setting. Ultrasonography (US) measurement is a more feasible and accessible method that can quantitatively and qualitatively evaluate muscles. Muscle thickness measured using US is similar to that measured using CT, MRI, or DEXA [8,9,10,11]. Therefore, US measurement has rapidly spread in the field of physical therapy.

The measurement of muscle thickness using US presents acceptable intra-rater reliability; however, there are a few reports on its inter-rater reliability [12,13,14]. Moreover, relative reliability cannot distinguish the type and extent of errors included in the measured values, as this statistical method is based on the assumption that only random errors occur. The measured error can be either systematic or random, and it is difficult to overcome systematic errors through repeated measurements. Hence, it is necessary to determine the presence of systematic errors.

Materials and methods

Aim and study design

This study aimed to investigate the intra- and inter-rater reliabilities and absolute reliability of measuring rectus femoris muscle thickness using US. The present study design was a test–retest study (within-day and 1 week apart).

Participants

A poster for research recruitment was posted in the first author's hospital to recruit participants for this study. Over a 12-month period (March 2017 to March 2018), 12 young male volunteers with a mean age of 26.5 ± 3.9 years (mean ± standard deviation) participated in this study. Participant characteristics are shown in Table 1. The inclusion criteria included: (1) over 20 years old; (2) under 40 years old; (3) no orthopedic problem; (4) not hypersensitive to electrical stimulation; (5) able to provide informed consent. All participants received an explanation of the study purpose and provided written informed consent. The study was approved by the ethics committees of Saitama Medical University International Medical Center (16–188).

Table 1 Demographics of the participants

Measurement of rectus femoris muscle thickness

In the measurements performed using US, a transducer probe (UF-850XTD: Fukuda Denshi Co. Ltd., Tokyo, Japan) was used, as illustrated in Fig. 1b. Participants lay in the supine position with the knee extended and maintained this posture throughout the US measurement (Fig. 1). Rectus femoris muscle thickness was measured from the right side of the thigh. The ultrasound probe was placed on the anterior surface of the thigh, at the midpoint of the length between the anterior superior iliac spine and the upper border of the patella, referring to the measurement position of Pardo et al. [14]. The measurement sites were determined using a measuring tape and marked with a permanent marker. A circumferential mark was made perpendicularly at the midpoint of the long axis. The probe was placed at the top of this circumferential line (Fig. 2) and moved along the line until a suitable image was obtained, in which a parallel position of the three borderlines of the rectus femoris, vastus intermedius, and femur was achieved (Fig. 3). On the screen of the US device, muscle thickness was calculated by setting the cursor to two points from the border lines of the rectus femoris to the borderlines of vastus intermedius. The probe was maintained in this perpendicular position on the skin surface, and minimal contact pressure was applied during the measurements to obtain good-quality images.

Fig. 1
figure 1

Ultrasonography set-up in the measurement of skeletal muscle thickness. a Participant is depicted in a supine position while the user holds the ultrasound probe against the anterior surface of the thigh. b The exact position of the probe is illustrated. Muscle thickness was determined from the resulting ultrasound images, and the values from two independent raters were used to decipher the intra- and inter-rater reliabilities

Fig. 2
figure 2

Ultrasound of rectus femoris muscle. Ultrasound images of the muscular layers were obtained, and the muscle thickness of the rectus femoris was measured by drawing a perpendicular line from the border lines of the rectus femoris to the border lines of the vastus intermedius. On the screen of the US (ultrasonography) device, muscle thickness was calculated by setting the cursor to two points from the border lines of the rectus femoris to the border lines of vastus intermedius. Top arrow: rectus femoris muscle; middle arrow: vastus intermedius muscle; bottom arrow: femur

Fig. 3
figure 3

Bland–Altman plots of intra- and inter-rater reliabilities of ultrasonography to measure skeletal muscle thickness. The values from two independent raters were used to decipher the intra- and inter-rater reliabilities. Bland–Altman plots of the mean measurement against the difference between the measurements are depicted for intra-rater reliability (a) and inter-rater reliability (b)

Rectus femoris muscle thickness was measured by two trained physical therapists (A and B). To investigate the intra-rater reliability, the measurement was performed twice with an interval of 1 week by only rater A. To investigate the inter-rater reliability, the measurement was concurrently performed by two independent raters (A and B); both raters were blinded to each other's results. We simultaneously focused on two or more measurements because we wanted to examine the inter-rater reliability by evaluating the concurrent measurement performed by two examiners.

Data analysis

The intraclass correlation coefficient (ICC) with 95% confidence intervals (CI) was calculated to determine the relative reliability. Inter- and intra-rater reliabilities were determined using the ICC (1, 1) and ICC (2, 1) methods, respectively. The absolute reliability was evaluated using Bland − Altman analysis for obtaining intra- and inter-rater reliabilities. Furthermore, we calculated a confidence level of 95% of the average difference between the two measured values to detect the existence of fixed errors, and this was performed using the following formula:

$$\stackrel{-}{d}\pm t\times \sqrt{\frac{\mathrm{SD}d}{n},}$$

where \(\stackrel{-}{d}\) is the average of 2 measurement values, t is the degree of freedom in the case of n−1, n is the sample size, and SDd is the standard error of \(\stackrel{-}{d}\)

If the result did not contain 0, we concluded that a fixed error was present, assuming that the measured values were distributed in a fixed direction [15]. Further, we calculated the T value to inspect for the existence of a proportional error using the following formula:

$$t=r\sqrt{\frac{n-2}{1-{r}^{2}};}$$

r is calculated using the correl function in Excel.

If the calculated T value was higher than the T value when the degree of freedom was n−2, and the level of significance was 5%, we concluded that a proportional error was present, assuming that there was a significant correlation [15]. Additionally, we calculated the minimal detectable change at the 95% level of confidence [MDC95 = standard error of measurement (SEM) × 1.96 × √2], which is determined from the standard deviation of the difference (SDd): SEM = SDd × √2. This represents the smallest change that can be interpreted as a real difference using the following formula:

$${\text{MDC}}_{{95}} \; = \;1.96\; \times \;{\text{SDd}}$$

Statistical analysis was performed using PASW Statistics ver. 18.0 (SPSS Japan Inc, Tokyo, Japan), with the level of significance set at 5%.

Results

The two measurements obtained by rater A were 16.7 ± 3.0 mm and 16.9 ± 3.3 mm, and the measurement obtained by rater B was 17.0 ± 2.3 mm (Table 2). According to the results of the Bland − Altman analysis, no fixed or proportional errors were present. The ICC (1, 1) for the intra-rater reliability was 0.95 (95% CI 0.856 − 0.987), and the ICC (2, 1) for the inter-rater reliability was 0.70 (95% CI 0.225−0.901). The MDC95 values of rectus femoris thickness for the intra-rater and inter-rater reliabilities were 2.0 mm and 4.3 mm, respectively (Table 3). The Bland−Altman plots were drawn (Fig. 4), and the differences in the scores were plotted against the mean scores of the measurements.

Table 2 Measurement values taken by 2 independent raters
Table 3 The relative and absolute reliability results

Discussion

In this study, we evaluated the intra- and inter-rater and absolute reliabilities of rectus femoris muscle measurement using US. Our findings support US as a useful tool for measuring rectus femoris muscle thickness in healthy young individuals. The primary use of analyses of healthy populations is to allow comparisons with pathological populations, and it follows, therefore, that measurement reliability in healthy populations is of importance. Few studies have compared MDC for intra- and inter-rater reliabilities; however, our study results indicated that rectus femoris muscle thickness should be measured by the same rater.

Based on the 95% CI Koo et al. [16] reported that ICC values less than 0.5, between 0.5 and 0.75, between 0.75 and 0.9, and greater than 0.9 indicate poor, moderate, good, and excellent reliability, respectively. The ICC values for the rectus femoris muscle thickness (intra-rater reliability = 0.99, inter-rater reliability = 0.96) presented here were similar to those reported by other studies [17, 18]. In this study, the ICCs of relative reliability was 0.95 for ICC (1,1) and 0.70 for ICC (2,1). ICC (1,1) and ICC (2,1), used to assess the intra- and inter-rater reliabilities, were at the “excellent” and “moderate” levels, respectively. Similarly, 95% CI was 0.856−0.987 for ICC (1,1) and 0.225−0.901 for ICC (2,1). ICC (1,1) had “good” to “excellent” reliability; however, ICC (2,1) significantly varied from “poor” to “excellent”. Therefore, the intra- rater reliability could be considered valid.

In addition, a systematic error was not detected in the absolute reliability results of the Bland−Altman analysis. The MDC95, a statistical parameter, can be used as a threshold to help users distinguish a real change from a measurement error for an individual participant. If the obtained value is equal to or less than the MDC95, then it is within the error range [19]. However, if the value is greater than the MDC95, it can be determined that a true change has occurred [19]. In this study, the MDC95 values for intra-rater and inter-rater reliabilities were found to be 2.0 mm and 4.3 mm, respectively, which is a useful index for judging the time course and therapeutic effect.

However, this study has some limitations. The amount of inward pressure on the patient caused by the transducer was not evaluated during the measurements. We believe that this is a possible reason why the inter-rater reliability was observed to be “moderate”. In other words, it is necessary to consider the difference in the rater’s measurement ability. These transducer conditions should have been precisely determined in the measurements obtained in the study. Another limitation is that only young males were enrolled in the study. Therefore, the influence of gender and age is unknown. Finally, a systematic error was not identified in this study. Thus, we should have calculated the appropriate repetition count to improve reliability. Further research is required to keep the amount of inward pressure caused by the transducer constant to ensure that inter-rater reliability is as good as intra-rater reliability. Similarly, more participants are needed to determine the effect of gender and age.

Conclusions

In conclusion, the intra-rater reliability and inter-rater reliability of the measurement of rectus femoris muscle thickness using US were mostly sufficient in healthy young men. Moreover, we determined the measurement error for rectus femoris thickness. Measuring the MDC is useful when verifying the degree of muscle atrophy and any therapeutic effects via the strengthening of muscles. Therefore, the measurement of rectus femoris muscle thickness using US could be considered applicable in clinical research.

Availability of date and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

CT:

Computed tomography

CI:

Confidence interval

DEXA:

Dual-energy X-ray absorptiometry

ICC:

Intraclass correlation coefficient

MRI:

Magnetic resonance imaging

MDC95 :

Minimal detectable change with 95% CI

SD:

Standard deviation

SEM:

Standard error of measurement

US:

Ultrasonography

References

  1. Lexell J, Taylor CC (1988) Sjöström M (1998) What is the cause of the ageing atrophy? Total number, size and proportion of different fiber types studied in whole vastus lateralis muscle from 15- to 83-year-old men. J Neurol Sci 84:275–294

    Article  CAS  Google Scholar 

  2. Evans WJ, Lexell J (1995) Human aging, muscle mass, and fiber type composition. J Gerontol A Biol Sci Med Sci 50:11–16

    Article  Google Scholar 

  3. Tsubahara A, Chino N, Akaboshi K, Okajima Y, Takahashi H (1995) Age-related changes of water and fat content in muscles estimated by magnetic resonance (MR) imaging. Disabil Rehabil 17:298–304

    Article  CAS  Google Scholar 

  4. Overend TJ, Cunningham DA, Paterson DH, Lefcoe MS (1992) Thigh composition in young and elderly men determined by computed tomography. Clin Physiol 12:629–640

    Article  CAS  Google Scholar 

  5. Frantzell A, Ingelmark BE (1951) Occurrence and distribution of fat in human muscles at various age levels; a morphologic and roentgenologic investigation. Acta Soc Med Ups 56:59–87

    CAS  PubMed  Google Scholar 

  6. DA YamadaY S, Nakamura E, Morimoto T, Kimura M, Oda S (2010) Extracellular water may mask actual muscle atrophy during aging. J Gerontol A Biol Sci Med Sci 65:510–516

    Article  Google Scholar 

  7. Wells J (2006) Validity of methods used for the assessment of body components in children. Horm Res 66:58–64

    CAS  Google Scholar 

  8. Jolley SE, Bunnell AE, Hough CL (2016) ICU-acquired weakness. Chest 150:1129–1140

    Article  Google Scholar 

  9. Kress JP, Hall JB (2014) ICU-acquired weakness and recovery from critical illness. N Engl J Med 370:1626–1635

    Article  CAS  Google Scholar 

  10. Thomaes T, Thomis M, Onkelinx S, Coudyzer W, Cornelissen V, Vanhees L (2012) Reliability and validity of the ultrasound technique to measure the rectus femoris muscle diameter in older CAD-patients. BMC Med Imag 12:7

    Article  Google Scholar 

  11. Dupont AC, Sauerbrei EE, Fenton PV, Shragge PC, Loeb GE, Richmond FJ (2001) Real-time sonography to estimate muscle thickness: comparison with MRI and CT. J Clin Ultrasound 29:230–236

    Article  CAS  Google Scholar 

  12. Vijay H, Gopi CK, RohitK AD, Saurabh M, Maroof AK, Karan M, Anant M, Randeep G (2017) Intra- and inter-observer reliability of quadriceps muscle thickness measured with bedside ultrasonography by clinical care physicians. Indian J Crit Care Med 21:448–452

    Article  Google Scholar 

  13. Silva CR, Costa AD, Rocha T, Lima DA, Nascimento TD, Moraes SR (2018) Quadriceps muscle architecture ultrasonography of individuals with type 2 diabetes: reliability and applicability. PLoS ONE 13:e0205724

    Article  Google Scholar 

  14. Pardo E, El Behi H, Boizeau P, Verdonk F, Alberti C, Lescot T (2018) Reliability of ultrasound measurements of quadriceps muscle thickness in critically ill patients. BMC Anesthesiol 18:205–212

    Article  Google Scholar 

  15. Ludbrook J (2002) Statistical techniques for comparing measurers methods of measurement: a critical review. Clin Exp Pharmacol Physiol 29:527–536

    Article  CAS  Google Scholar 

  16. Koo TK, Li MY (2016) A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr 15:155–163

    Article  Google Scholar 

  17. Thoirs K, English C (2009) Ultrasound measures of muscle thickness: intra- examine reliability and influence of body position. Clin Physiol Funct Imag 29:440–446

    Article  Google Scholar 

  18. Tillquist M, Kutsogiannis DJ, Wischmeyer PE, Kummerlen C, Leung R, Stollery D, Karvellas CJ, Preiser JC, Bird N, Kozar R, Heyland DK (2014) Bedside ultrasound is a practical and reliable measurement tool for assessing quadriceps muscle layer thickness. J Parenter Enter Nutr 38:886–890

    Article  Google Scholar 

  19. de Vet HC, Terwee CB, Knol DL, Bouter LM (2006) When to use agreement versus reliability measures. J Clinical Epidemiol 59:1033–1039

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This research received no specific funding.

Author information

Authors and Affiliations

Authors

Contributions

YF designed the study. YT, YF and KM performed the data collection. YT and YF analyzed and interpreted the data and drafted the manuscript. AT contributed how to use the ultrasonography. TM reviewed and revised the manuscript. SM provided medical advice and overall study management. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Yuji Fujino.

Ethics declarations

Ethics approval and consent to participate

All participants provided written informed consent, and the study was approved by the ethics committees of Saitama Medical University International Medical Center (16–188).

Consent for publication

The participants consented to the publication of their images.

Competing interests

Shigeru Makita received honorarium from Otsuka Pharmaceutical Co., Ltd. and Daiichi Sankyo Co., Ltd. and Fukuda Electronics Co., Ltd.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Takahashi, Y., Fujino, Y., Miura, K. et al. Intra- and inter-rater reliability of rectus femoris muscle thickness measured using ultrasonography in healthy individuals. Ultrasound J 13, 21 (2021). https://doi.org/10.1186/s13089-021-00224-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13089-021-00224-8

Keywords