Skip to main content

Table 2 Intra-researcher variability for each POCUS parameter

From: Intra-and inter-observer variability of point of care ultrasound measurements to evaluate hemodynamic parameters in healthy volunteers

Parameter

Intra-observer variability based on mean CV in %

Observer 1 (N = 540)

Observer 2 (N = 469)

Observer 3 (N = 493)

Mean (N = 1502)

Measured

 LVOT-D

2.54

2.96

2.18

2.56

 LVOT-VTI

5.66

6.22

7.34

6.41

 EPSS

23.07

19.74

34.25

25.69

 MAPSE

6.25

10.46

9.97

8.89

 TAPSE

4.86

8.26

6.18

6.43

 IVC-D

4.45

4.22

4.82

4.50

 rCBF-D

4.91

7.81

4.82

5.85

 rCBF-VTI

10.09

10.72

10.89

10.57

Calculated

 CO

5.48

6.24

7.26

6.33

 SV

5.50

6.01

7.68

6.39

 IVC-CI

12.25

19.63

18.46

16.78

 rCBF

11.63

18.91

14.03

14.86

  1. CV coefficient of variation, LVOT-D left ventricular outflow tract diameter, LVOT VTI left ventricular outflow tract velocity time integral, EPSS E point septal separation, MAPSE mitral annular plane systolic excursion, TAPSE tricuspid annulus plane systolic excursion, IVC-D inferior vena cava diameter, rCBF-D right Carotid blood flow diameter, rCBF-VTI right Carotid blood flow velocity–time integral, CO Cardiac output, SV stroke volume, IVC-CI inferior vena cava collapsibility index, rCBF right Carotid blood flow